The Hough Transform – An Amazing Tool.

Shawn Wallace, Applications Engineer, EDAX

Part of my job is understanding and pushing the limits of each part of our systems. One of the most fundamental parts of the EBSD system is the Hough Transform. The Hough Transform role is finding the lines on an EBSD pattern. This is the first step in indexing a pattern (Fig. 1). If this step is not consistent, the quality of any indexing and any derivative data is questionable. A normal user does not really need to understand all the intricacies of every part of the system, but it still is worthwhile to understand how your data and data quality can be affected.

Figure 1: On the left are the overlaid lines found via the Hough Transform. On the right is the Indexed solution overlaid based on the Hough. The quality of the indexed solution is based on the quality of the Hough.

With that in mind, I ran an experiment on a steel sample to see how far the Hough could be pushed and still give consistent indexing. For this experiment, I used our Hikari Super at a series of different binnings between its native resolution of 640X480 Pixels at 1×1 binning down to 35×26 pixels at 18×18 binning. All pixel resolutions are noted in Table 1. I kept my Hough Settings and beam settings consistent. My only other variable was exposure to get the camera to be equally saturated at around 0.85 saturation.

I expected the lower binning Patterns to be consistent and they were (Fig. 2). All three Euler Angles between the 1×1, 2×2, 4×4, and 8×8, were within 0.4 degrees of each other. Pushing the camera and the Hough even further really surprised me though.

Figure 2: Indexed Pattern for the lower binning showed a remarkable consistency in indexing.

Figure 3: The indexing results still held their consistency even for highest binning settings used.

I expected some drop off with the consistency of the orientation when I dropped my binning to 10×10, 16×16, and even 18×18 and it did not fully materialize (Fig. 3). The range did broaden in the Euler Angles, specifically ᶲ₂’s range increased to 3 degrees, but that is change of <1% given the entire range for ᶲ₂ is 360 degrees. Table 1 shows the data is the raw form. Overall, the data is great, from low to high binning with minimal loss in in our indexing metrics (CI and Fit) and consistency in Euler Angles except for the 18×18 binning. That is where we have found our limit, specifically when it comes to indexing metrics. We see a sharp drop off in the CI. The pixilation of the pattern has gotten to a point where it is difficult to find a unique solution. This drop off is why we tell our customer that 16×16 is the limit of binning they should use for reliable, high quality data.

Table 1. Indexing Metrics and Euler Angles for all data points.

With all that said, most EBSD work is not on a single orientation, but a map. Does this hold true on a map? It does. In Figure 4 and Figure 5, we can see the mapping results for 2×2 binning and 10×10 binning. Both indexed at 99.9% with their average CI’s being 0.89 and 0.84 respectively, with very little change in orientations. This level of data quality across binnings is why EDAX uses the Hough. It is an amazing little tool.

Figure 4. This map was taken at 2×2 binning. Internal deformation of the grains is visible, with inclusions between relatively undeformed.

Figure 5. This map was taken at 10×10 binning in approximately the same area as Figure 4. Again, internal deformation is showed in the larger grain, while the inclusions are undeformed.

Considerations for your New Year’s Resolutions from Dr. Pat

Dr. Patrick Camus, Director of Research and Innovation, EDAX

The beginning of the new calendar year is a time to reflect and evaluate important items in your life. At work, it might also be a time to evaluate the age and capabilities of the technical equipment in your lab. If you are a lucky employee, you may work in a newly refurbished lab where most of your equipment is less than 3 years old. If you are such a fortunate worker, the other colleagues in the field will be envious. They usually have equipment that is much more than 5 years old, some of it possibly dating from the last century!

Old Jalopy circa 1970 EDAX windowless Si(Li) detector circa early 70’s

In my case, at home my phone is 3 years old and my 3 vehicles are 18, 16, and 3 years old. We are definitely evaluating the household budget this year to upgrade the oldest automobile. We need to decide what are the highest priority items and which are not so important for our usage. It’s often important to sort through the different features offered and decide what’s most relevant … whether that’s at home or in the lab.

Octane Elite Silicon Drift Detector 2017 Dr. Pat’s Possible New Vehicle 2017

If your lab equipment is older than your vehicles, you need to determine whether the latest generation of equipment will improve either your throughput or the quality of your work. The latest generations of EDAX equipment can enormously speed up throughput and the improve quality of your analysis over that of previous generations – it’s just a matter of convincing your boss that this has value for the company. There is no time like the present for you to gather your arguments into a proposal to get the budget for the new generation of equipment that will benefit both you and the company.
Best of luck in the new year!